Repair of Breakdown of high voltage cable assembly - Newheek X ray Cable - Newheek X ray Cable

Repair of Breakdown of high voltage cable assembly

The breakdown problem of x ray high voltage cable is very common. Next, I will introduce the repair methods of breakdown of high voltage cable assembly.
Repair of Breakdown of high voltage cable assembly
Generally, there are two common breakdown modes. One is the reason that the rubber layer of high voltage cable assembly head does not bond well when pressed, and there is a large air gap between layers. The second is that the part of the high voltage cable head directly breaks through the rubber cone head.
The arc holes can be seen on the surface of the high voltage cable assembly head in both cases, which need to be handled well. When removing carbides, the breakdown site should be incised and expanded, while incising and observing, and all the damaged parts should be exposed to avoid adverse consequences. After removing carbides, the carbides are treated according to the above cleaning methods and filled with one-component vulcanized silicone rubber or epoxy resin E44 as appropriate, with a certain amount of external force, so as to bond reliably. In the process of repairing the breakdown part, careful treatment should be taken so that the air gap can not be left to adhere reliably.
Repair of Breakdown of high voltage cable assembly


(+86) 18953679166
service@newheek.com